Congenital: Fontan| Volume 158, ISSUE 6, P1627-1636, December 2019

Cavopulmonary assist: Long-term reversal of the Fontan paradox

Open ArchivePublished:August 28, 2019DOI:https://doi.org/10.1016/j.jtcvs.2019.06.112

      Abstract

      Objective

      Fontan circulatory inefficiency can be addressed by replacing the missing subpulmonary power source to reverse the Fontan paradox. An implantable cavopulmonary assist device is described that will simultaneously reduce systemic venous pressure and increase pulmonary arterial pressure, improving preload and cardiac output, in a univentricular Fontan circulation on a long-term basis.

      Methods

      A rotary blood pump that was based on the von Karman viscous pump was designed for implantation into the total cavopulmonary connection (TCPC). It will impart modest pressure energy to augment Fontan flow without risk of obstruction. In the event of rotational failure, it is designed to default to a passive flow diverter. Pressure-flow performance was characterized in vitro in a Fontan mock circulatory loop with blood analog.

      Results

      The pump performed through the fully specified operating range, augmenting flow in all 4 directions of the TCPC. Pressure rise of 6 to 8 mm Hg was readily achieved, ranging to 14 mm Hg at highest speed (5600 rpm). Performance was consistent across a wide range of cardiac outputs. In stalled condition (0 rpm), there was no discernible pressure loss across the TCPC.

      Conclusions

      A blood pump technology is described that can reverse the Fontan paradox and may permit a surgical strategy of long-term biventricular maintenance of a univentricular Fontan circulation. The technology is intended for Fontan failure in which right-sided circulatory inefficiencies predominate and ventricular systolic function is preserved. It may also apply before clinical Fontan failure as health maintenance to preempt the progression of Fontan disease.

      Key Words

      Abbreviations and Acronyms:

      MCS (mechanical circulatory support), TCPC (total cavopulmonary connection)
      Figure thumbnail fx1
      Cavopulmonary assist device: flow augmentation through the total cavopulmonary connection.
      An implantable blood pump that can safely and reliably reverse the Fontan paradox for the long term may enable a physiologically curative strategy for single-ventricle heart disease.
      Fontan palliation is associated with a chronic circulatory inefficiency that culminates in secondary disease, failure, and attrition. We report the feasibility of an implantable cavopulmonary assist device that can safely replace the missing subpulmonary power source and reverse the Fontan paradox for the long term. This may enable a physiologically curative strategy for single functional ventricle.
      See Commentaries on pages 1637, 1639, and 1641.
      Repair of single functional ventricle currently consists of a staged surgical approach that terminates in a univentricular Fontan circulation. Since its introduction, Fontan palliation has dramatically improved the outcome for all forms of single-ventricle heart disease
      • Fontan F.
      • Baudet E.
      Surgical repair of tricuspid atresia.
      • Kreutzer G.
      • Galindez E.
      • Bono H.
      • DePalma C.
      • Laura J.P.
      An operation for the correction of tricuspid atresia.
      ; however, survivors of the Fontan procedure are committed for the remainder of their lives to a state of chronic circulatory inefficiency because of the lack of a subpulmonary ventricle. This inefficiency results in coexisting elevated systemic venous pressure and reduced cardiac output, known clinically as the Fontan paradox.
      • de Leval M.R.
      The Fontan circulation: what have we learned? What to expect?.
      Combined, these problems have been clearly shown to progress to late Fontan failure and attrition.
      • Rychik J.
      The relentless effects of the Fontan paradox.
      Despite extensive clinical and research efforts, therapies to prevent or reverse this circulatory decline remain extremely limited, with no breakthroughs for decades.
      • Gewillig M.
      • Brown S.C.
      The Fontan circulation after 45 years: update in physiology.
      With an increasing number of late survivors of the Fontan operation, Fontan failure is now a routinely encountered and challenging clinical problem.
      We have theorized that replacement of the missing subpulmonary power source would reverse the Fontan paradox and shift the univentricular circulation toward more stable biventricular equivalency.
      • Rodefeld M.D.
      • Boyd J.H.
      • Myers C.D.
      • Lalone B.J.
      • Bezruczko A.
      • Potter A.W.
      • et al.
      Cavopulmonary assist: circulatory support for the univentricular Fontan circulation.
      It would decongest the systemic venous and lymphatic circulations, while simultaneously improving cardiac output. If applied as durable long-term support, it could in theory preempt Fontan-associated disease progression by maintaining biventricular equivalency.
      Despite the simplicity and attractiveness of this concept, the technologic considerations are complex and dissimilar to any other mechanical circulatory support (MCS) application. We have found that existing MCS technologies are ill-suited and unmodifiable to address Fontan circulatory inefficiency. Alternatively, we have focused on the development of an anatomically specific device that can replace the missing subpulmonary power source and reverse the Fontan paradox.
      • Rodefeld M.D.
      • Coats B.
      • Fisher T.
      • Giridharan G.A.
      • Chen J.
      • Brown J.W.
      • et al.
      Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
      This report describes the computational and in vitro basis for an implantable cavopulmonary assist device concept that has potential to restore biventricular equivalency safely and durably in a univentricular circulation.

      Methods

      Pump Design

      In previous studies, we have demonstrated that a remotely powered rotary pump, which was based on the von Karman viscous pump, can provide 4-way flow augmentation through the total cavopulmonary connection (TCPC).
      • Rodefeld M.D.
      • Coats B.
      • Fisher T.
      • Giridharan G.A.
      • Chen J.
      • Brown J.W.
      • et al.
      Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
      A biconical impeller suspended in the midst of the TCPC functions as a 2-sided centrifugal pump, drawing inflow from the superior and inferior venae cavae and augmenting outflow to the left and right pulmonary arteries at physiologic pressure. A modest pressure step-up of 6 mm Hg has been shown in vitro and in vivo to restore the univentricular Fontan circulation to biventricular equivalency under normal physiologic conditions.
      • Rodefeld M.D.
      • Boyd J.H.
      • Myers C.D.
      • Lalone B.J.
      • Bezruczko A.
      • Potter A.W.
      • et al.
      Cavopulmonary assist: circulatory support for the univentricular Fontan circulation.
      • Rodefeld M.D.
      • Coats B.
      • Fisher T.
      • Giridharan G.A.
      • Chen J.
      • Brown J.W.
      • et al.
      Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
      We have also demonstrated that a cavopulmonary assist device that is based on the von Karman viscous pump will not obstruct the systemic venous pathway in the event of rotational dysfunction and will instead reduce kinetic energy loss in the TCPC.
      • Rodefeld M.D.
      • Coats B.
      • Fisher T.
      • Giridharan G.A.
      • Chen J.
      • Brown J.W.
      • et al.
      Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
      Because an external power source would limit such a device to temporary use, a permanent power source was incorporated in the form of an outrunner brushless direct current motor. In an outrunner motor, the rotating component is outside rather than inside (Figure 1). For an in-series pump situated in the Fontan venous pathway, this configuration is key to eliminating the risk of venous pathway obstruction because it allows preservation of a wide passage between the impeller and housing. As a result, the pump can serve in 2 distinctly different capacities: rotating, it functions as a pump; non-rotating, it functions as a static flow diverter (Figure 2). Thus, it is intended to be beneficial even if it fails.
      Figure thumbnail gr1
      Figure 1Outrunner brushless direct current motor. A, Traditional motor design with inner rotating component fabricated with permanent magnets driven by the outer, stationary motor windings, depicted in orange. B, Outrunner motor design with outer rotating component with permanent magnets driven by stationary central motor windings, depicted in orange. The outer rotating component is a critical innovation that enables the cavopulmonary assist device to be implanted within the total cavopulmonary connection.
      Figure thumbnail gr2
      Figure 2Implantable cavopulmonary assist device. A, Conceptual image. A central impeller contains the motor (impeller support shaft and struts not shown). The external housing is a thin-walled, hard case shell that has no component bulk and serves only to direct Fontan flow. The wide gap between the impeller and the housing must support passive total cavopulmonary connection flow at low resistance in the event of device dysfunction. B, Advanced prototype image depicting pump function. Color contrast demonstrates superior vena caval (SVC) flow (red) and inferior vena caval (IVC) flow (blue) mixing to both left pulmonary arterial (LPA) and right pulmonary arterial (RPA) outlets (purple).
      A biconical stator, containing iron laminations and copper windings, is located centrally within the impeller, and has no blood contact or moving parts. It is surrounded by a biconical rotor with embedded magnets as the only moving component. Rotor surface vanes provide low-pressure, high-volume augmentation of cavopulmonary flow (Video 1). A secondary (internal) flow path between the rotor and the stator is an inherent centrifugal pump and integrates several key functions: (1) electromagnetic flux transfer for motor torque, (2) heat dissipation, (3) bearing lubrication, and (4) rotodynamic stabilization (Video 2). The housing is a double-inlet, double-outlet, thin-walled, hard case shell that serves only to direct TCPC flow and structurally suspend the pump. Wiring from the stator travels through the central shaft to exit the housing via shaft struts. Because there is no external component bulk, the pump is no larger than the native TCPC and is suitable for in situ implantation.

      Pump Fabrication and Testing

      An advanced prototype was designed and fabricated at adult scale (25 mm diameter inlets, 20 mm diameter outlets; Figure 3). With the aid of computational fluid dynamic modeling, impeller dimension and motor size were determined on the basis of the torque required to achieve a pressure rise of 6 to 8 mm Hg under physiologic conditions. Noncircumferential conical bearings support the rotor at each end. The impeller surface vane height is only 0.7 mm. The pump was tested initially in an open circuit, with no baseline imposed flow, with blood analog (water/glycerin 63:37, physiologic density 1060 kg/m3, 3.2 cP). Testing was then performed in a static loop for pressure-flow performance and finally in vitro in a mock loop of a Fontan circulation for physiologic performance. For static performance, the pump was operated at 0 to 5600 rpm against 5 different resistances. Steady-state pressure head and flow rates were plotted to characterize hydraulic (pressure-flow) performance. Inlet partial and complete occlusion performances were assessed. For physiologic performance testing, a Fontan mock circulatory system was used that includes compliance and resistance elements.
      • Rodefeld M.D.
      • Coats B.
      • Fisher T.
      • Giridharan G.A.
      • Chen J.
      • Brown J.W.
      • et al.
      Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
      Figure thumbnail gr3
      Figure 3Advanced prototype, adult scale. A, Design solid model showing housing, support struts, and motor, with mounted impeller. Wiring travels from the stator through the central shaft and strut to exit the housing. A thermocouple embedded in the stator core was used to monitor device operating temperature. The intrinsic heat-sink properties of the internal flow path result in excellent heat performance, as observed by a decrease in core temperature at higher speeds. The only moving component is the rotor. A secondary flow path allows for inflow at each axial end, with outflow at the equatorial gap (). B, Design schematic showing patulous channel to prevent obstruction in the event of rotational failure. Rotor surface shows 0.7 mm height curved surface vanes. C, Assembled prototype as tested and demonstrated in the videos.

      Results

      All design objectives were met, including the following: (1) pressure-flow performance of 0 to 14 mm Hg pressure rise at 0 to 5600 rpm; (2) mean scalar shear stress less than 100 Pa, transit time less than 0.01 seconds; (3) noncavitating performance; (4) acceptable recirculation; and (5) no pressure loss in the static condition (0 rpm). Heat dissipation was excellent, confirmed by reduction of core temperature at higher speed and no measurable increase in blood analog temperature (<0.1°C). Hydraulic performance was consistent across a wide range of cardiac output (Figure 4). A peak pressure of 14 mm Hg was achieved at the highest speed. The pump generated 5 L/min of flow and 6 mm Hg pressure head at 4000 rpm, and 5 L/min of flow and 8 mm Hg pressure head at 5600 rpm. The in vitro pump performance was consistent with the design specifications and computational fluid dynamic modeling (6-8 mm Hg pressure rise under physiologic conditions). The pump generated inflow from both the superior and inferior venae cavae equally to both pulmonary arteries for superior to inferior vena caval inflow ratios ranging from 30:70 to 50:50. The pump effectively mixed inflows from both inlets, with consistently symmetric outflow distribution (of importance to hepatic factor distribution; Video 1). The power draw was less than 6 W at maximal speed, and this is expected to improve with further optimization. Because the surface vane profile is so low, there is flexibility to increase vane height for higher pressure performance without compromising the no-obstruction constraint.
      Figure thumbnail gr4
      Figure 4Hydraulic performance was tested in a static mock flow loop with 2 inlets, 2 outlets, a storage chamber, and variable outflow resistance. The pressure head generated was plotted against pump-generated flow for multiple levels of outflow resistance for various pump speed settings. The pressure-flow curve demonstrates that the cavopulmonary assist device is capable of producing as much as 14 mm Hg pressure head.
      In in vitro physiologic mock loop testing, the device effectively reversed the Fontan (Table 1). With increasing rpm, ventricular end-diastolic pressure and aortic pressure increased. At 5000 rpm, the pump converted a cavopulmonary pressure head of positive 2.8 mm Hg to negative 4.8 mm Hg, replacing a small energy loss with a proportionally larger gain. It boosted pulmonary arterial pressure by 6 mm Hg and increased preload sufficient to increase cardiac output by 12.8%. Vena caval pressure was reduced 1.4 mm Hg, which is significant in this segment of the circulation, where even small reduction of pressure is beneficial to splanchnic health. Increase in pump speed beyond 5000 rpm caused a larger negative cavopulmonary pressure head but did not lead to a measurable increase cardiac output. When the pump was stopped (0 rpm), there was no discernible pressure drop, validating the no-obstruction design constraint. Importantly, in the stopped condition, TCPC turbulence was reduced relative to the condition with no pump installed.
      Table 1In vitro mock loop testing demonstrating reversal of the failing Fontan circulation with cavopulmonary assist
      CaseVIP speed (rpm)HR (beats/min)CO (L/min)CO % increaseVCP (mm Hg)PAP (mm Hg)CPPH (mm Hg)AoP (mm Hg)
      Fontan baseline0753.917.514.72.898.3/57.4
      Fontan + VIP3000754.15.1%1716.60.4102.3/60.5
      Fontan + VIP4000754.310.3%16.818−1.2111.9/66.7
      Fontan + VIP5000754.412.8%16.120.8−4.8113.6/69.0
      At 5000 rpm, the pump converted a cavopulmonary pressure head of positive 2.8 mm Hg to negative 4.8 mm Hg, replacing a small energy loss with a proportionally larger gain. It boosted pulmonary arterial pressure by 6 mm Hg and increased preload sufficient to increase cardiac output by 12.8%, demonstrating reversal of a failing Fontan circulation. Computational modeling and in vitro mock loop studies demonstrate consistent and symmetric outflow to the right and left pulmonary arteries. In clinical use, outflow distribution is expected to be affected by patient-specific pulmonary resistance. Fontan baseline includes pump in total cavopulmonary connection in stalled (0 rpm) condition. VIP, Viscous impeller pump; HR, heart rate; CO, cardiac output; CO % increase, percentage increase from baseline Fontan cardiac output; VCP, vena caval pressure; PAP, pulmonary artery pressure; CPPH, cavopulmonary pressure head (pump ΔP = vena caval pressure − pulmonary artery pressure); AoP, aortic pressure.

      Discussion

      Among the many significant advances in the treatment of congenital heart disease during the past 5 decades, single-ventricle palliation has had perhaps the largest effect. Nonetheless, it was predicted by its innovators that it may be incompatible with a normal lifespan and require improvement or modification.
      • Fontan F.
      • Kirklin J.W.
      • Fernandez G.
      • Costa F.
      • Naftel D.C.
      • Tritto F.
      • et al.
      Outcome after a “perfect” Fontan operation.
      • Kreutzer C.
      • Kreutzer J.
      • Kreutzer G.O.
      Reflections on five decades of the Fontan Kreutzer procedure.
      • de Leval M.R.
      The Fontan circulation: a challenge to William Harvey?.
      As predicted, Fontan-associated disease now represents one of the most significant challenges in the field. Although Fontan palliation is lifesaving, it carries a lifelong chronic disease burden that has no primary preventive therapy and is ultimately life limiting.
      • Rychik J.
      Forty years after the Fontan: a failed strategy.
      • Veldtman G.R.
      • Opotowsky A.R.
      • Wittekind S.G.
      • Rychik J.
      • Penny D.J.
      • Fogel M.
      • et al.
      Cardiovascular adaptation to the Fontan circulation.
      As a reflection of its palliative nature, late Fontan attrition is relatively constant, suggesting that immutable factors play a role, and 30-year survival is only 43% to 70%.
      • Diller G.P.
      • Kempny A.
      • Alonso-Gonzalez R.
      • Swan L.
      • Uebing A.
      • Li W.
      • et al.
      Survival prospects and circumstances of death in contemporary adult congenital heart disease patients under follow-up at a large tertiary centre.
      • Pundi K.N.
      • Johnson J.N.
      • Dearani J.A.
      • Li Z.
      • Hinck C.A.
      • Dahl S.H.
      • et al.
      40-year follow-up after the Fontan operation: long-term outcomes of 1,052 patients.
      • Poh C.L.
      • Zannino D.
      • Weintraub R.G.
      • Winlaw D.S.
      • Grigg L.E.
      • Cordina R.
      • et al.
      Three decades later: the fate of the population of patients who underwent the atriopulmonary Fontan procedure.
      Solutions to address Fontan-associated disease have been largely incremental and of unclear benefit, and we may have reached a plateau. Fontan optimization can at best only prolong an inherently inefficient system, because it does not address the underlying physiologic deficit: the lack of a subpulmonary ventricle. To date, little effort has focused on mechanical replacement of a subpulmonary power source, presumably because a comprehensive solution is both not obvious and highly complex. Replacement of the missing subpulmonary power source is, however, physiologically compelling; it will normalize systemic venous pressure, decongest the lymphatic circulation, and normalize ventricular loading conditions, emulating biventricular efficiency. Given the ongoing problems with Fontan-associated disease, it is reasonable to reconsider the current paradigm. Rather than Fontan perpetuation, a preferable strategy may be to reverse the Fontan.
      • Rodefeld M.D.
      The Fontan circulation: time for a moon shot?.

      Fontan Failure Is Largely Nonventricular Failure

      Although patients with failing Fontan circulations may exhibit classic features of congestive heart failure, the primary cause is not typically ventricular systolic dysfunction. Before end-stage Fontan disease, ejection fraction is preserved in more than 70% of patients; diastolic dysfunction predominates.
      • Anderson P.A.
      • Sleeper L.A.
      • Mahony L.
      • Colan S.D.
      • Atz A.M.
      • Breitbart R.E.
      • et al.
      Contemporary outcomes after the Fontan procedure: a pediatric heart network multicenter study.
      Although the mechanisms are complex and multifactorial, Fontan diastolic dysfunction has been largely attributed to chronic preload deprivation.
      • Averin K.
      • Hirsch R.
      • Seckeler M.D.
      • Whiteside W.
      • Beekman R.H.
      • Goldstein B.H.
      Diagnosis of occult diastolic dysfunction late after the Fontan procedure using a rapid volume expansion technique.
      • De Mey W.
      • Cools B.
      • Heying R.
      • Budts W.
      • Louw J.J.
      • Boshoff D.E.
      • et al.
      Can a volume challenge pinpoint the limiting factor in a Fontan circulation?.
      With time, abnormal loading conditions (decreased preload, increased afterload) result in cardiomyopathic remodeling and fibrosis,
      • Rathod R.H.
      • Prakash A.
      • Powell A.J.
      • Geva T.
      Myocardial fibrosis identified by cardiac magnetic resonance late gadolinium enhancement is associated with adverse ventricular mechanics and ventricular tachycardia late after Fontan operation.
      with eventual systolic dysfunction. It thus can be said that the Fontan circulation predisposes the single ventricle to fail. Furthermore, efforts that target improvement in ventricular function as a solution to Fontan disease may be misguided; it may be of greater benefit to provide cavopulmonary assistance.
      Cavopulmonary assist specifically addresses the lack of a subpulmonary power source, and it will in turn normalize ventricular loading conditions and improve function. In the setting of diastolic dysfunction but preserved systolic function, a modest increase in preload (approximately 1-3 mm Hg) will improve myocardial performance and cardiac output (opposite primary myocardial failure).
      Transplantation is, in a sense, a definitive therapy for Fontan failure, in that it reinstates a subpulmonary ventricle, but it does so at the cost of a different disease process. There are not sufficient donor organs available to provide transplants for all patients with single ventricle disease. Because of listing criteria, most patients with Fontan circulations will not receive a donor organ until they have end-stage disease with advanced comorbidities. Transplant survival of less than 50% at 10 years is arguably no better than the natural history of late Fontan attrition.
      • Ross H.J.
      • Law Y.
      • Book W.M.
      • Broberg C.S.
      • Burchill L.
      • Cecchin F.
      • et al.
      American Heart Association Adults With Congenital Heart Disease Committee of the Council on Clinical Cardiology and Council on Cardiovascular Disease in the Young, the Council on Cardiovascular Radiology and Intervention, and the Council on Functional Genomics and Translational Biology
      Transplantation and mechanical circulatory support in congenital heart disease: a scientific statement from the American Heart Association.
      For these reasons, transplantation is not a comprehensive solution and represents end-stage therapy.

      Fontan Optimization Falls Short

      Strategies to optimize Fontan hemodynamics have logically targeted the TCPC, where anatomic factors may result in significant power loss.
      • Bove E.L.
      • de Leval M.R.
      • Migliavacca F.
      • Balossino R.
      • Dubini G.
      Toward optimal hemodynamics: computer modeling of the Fontan circuit.
      • Sundareswaran K.S.
      • Pekkan K.
      • Dasi L.P.
      • Whitehead K.
      • Sharma S.
      • Kanter K.R.
      • et al.
      The total cavopulmonary connection resistance: a significant impact on single ventricle hemodynamics at rest and exercise.
      • Tang E.
      • Restrepo M.
      • Haggerty C.M.
      • Mirabella L.
      • Bethel J.
      • Whitehead K.K.
      • et al.
      Geometric characterization of patient-specific total cavopulmonary connections and its relationship to hemodynamics.
      These have included various vena caval offset configurations, as well as a Fontan Y conduit to reduce power loss in the inferior vena caval distribution.
      • Marsden A.L.
      • Bernstein A.J.
      • Reddy V.M.
      • Shadden S.C.
      • Spilker R.L.
      • Chan F.P.
      • et al.
      Evaluation of a novel Y-shaped extracardiac Fontan baffle using computational fluid dynamics.
      Although these modifications have been applied clinically, they have not had significant effect. The maximal pressure gain that can be derived from passive flow optimization (approximately 1-2 mm Hg) does not fully correct the circulatory deficit and, as a result, does not meaningfully improve circulatory status. By comparison, consider normal biventricular physiology, in which the right ventricle provides a cavopulmonary pressure step-up of approximately 6 to 8 mm Hg. To make a clinically meaningful improvement in Fontan, it is necessary to add pressure energy at the TCPC by a similar amount.

      Limitations of Current MCS Technology Applied to Fontan Failure

      The application of existing MCS devices for Fontan failure is limited, highly variable, and associated with poor outcomes.
      • Arnaoutakis G.J.
      • Blitzer D.
      • Fuller S.
      • Eckhauser A.W.
      • Montenegro L.M.
      • Rossano J.W.
      • et al.
      Mechanical circulatory support as bridge to transplantation for the failing single ventricle.
      • Horne D.
      • Conway J.
      • Rebeyka I.M.
      • Buchholz H.
      Mechanical circulatory support in univentricular hearts: current management.
      • Adachi I.
      • Burki S.
      • Fraser Jr., C.D.
      Current status of pediatric ventricular assist device support.
      • Woods R.K.
      • Ghanayem N.S.
      • Mitchell M.E.
      • Kindel S.
      • Niebler R.A.
      Mechanical circulatory support of the Fontan patient.
      It has generally focused on the application of systemic support in end-stage Fontan disease because of the nature of existing technologies. Nearly all MCS technologies are intended for systemic support. Depending on the circumstances, however, the device is likely mismatched to Fontan circulatory needs, that is, left-sided support to address a right-sided deficit. Anecdotally, existing MCS appears to be better suited for ventricular “pump” failure in Fontan circulations, rather than for Fontan failure secondary to lack of a subpulmonary ventricle. In the setting of preserved systolic function, systemic MCS in a Fontan circulation is redundant; it may needlessly congest the right-sided circulation where the circulatory bottleneck exists and exacerbate Fontan disease.
      Existing MCS applied strategically to the right side of the Fontan circulation has been reported
      • Prêtre R.
      • Häussler A.
      • Bettex D.
      • Genoni M.
      Right-sided univentricular cardiac assistance in a failing Fontan circulation.
      ; however, it is logistically impractical for several reasons: (1) the Fontan must be taken down to accommodate a single-inflow, single-outflow device; (2) a systemic MCS device is not optimized to the very low pressure rise desired for cavopulmonary assist, increasing risk for device failure; (3) the right-sided circulation is dependent on the device, making device dysfunction lethal; and (4) device operational lifespan is limited and therefore not a long-term solution. Given these considerations, the use of existing MCS technology in Fontan circulations is limited to end-stage disease as a bridge therapy, if it is used at all.

      Technical Considerations for Long-Term Fontan Reversal

      Dedicated right-sided circulatory support of the Fontan circulation presents a number of caveats that are uniquely challenging (Table 2). Technically, a cavopulmonary assist device is not a ventricular assist device, because it provides support where no ventricle exists. It is intended to function as a low-input auxiliary right-sided pump to maintain low systemic venous pressure and preload to the systemic ventricle, identical to the essential function of the right ventricle in a biventricular circulation.
      • Furey S.A.
      • Zieske H.A.
      • Levy M.N.
      The essential function of the right ventricle.
      Table 2Desirable technical features for a long-term cavopulmonary assist device
      • 1.
        Simplicity: single pump, with 1 moving part, that serves as a low-input primer for the systemic ventricle.
      • 2.
        Multidirectional flow: flow augmentation in all 4 distributions of the TCPC (SVC/IVC inflow, RPA/LPA outflow).
      • 3.
        Low-pressure performance: low-pressure (approximately 6-8 mm Hg), high-volume flow augmentation, similar to normal right ventricular hemodynamics.
      • 4.
        Wide performance range: consistent pressure rise across a wide range of cardiac outputs and variable physiologic demands, independent of pump speed.
      • 5.
        Nonobstructive: an inline pump in the Fontan venous pathway must account for pump failure, with no obstruction risk as a failure default.
      • 6.
        TCPC as the preferred default flow path: the failed (0 rpm) condition should ideally optimize passive TCPC flow as an unsupported Fontan circulation.
      • 7.
        Permissive preload and afterload performance: low hydraulic efficiency to reduce risks of both excessive negative upstream pressure (suction, vein collapse, cavitation) and excessive positive downstream pressure (perfusion lung injury). It will permit tolerance to the inflow instabilities of systemic venous return (supine vs upright posture, Valsalva).
      • 8.
        Mixing: symmetric hepatic factor distribution to prevent pulmonary arteriovenous malformation (Video 1).
        Figure thumbnail fx2
        Video 1Primary (external) flow path. With an open static circuit, flow is induced by the pump. Yellow and blue contrast represent systemic venous inflow. The green outflow represents left and right pulmonary arterial flow. Mixing indicates symmetric hepatic factor distribution. Video available at: https://www.jtcvs.org/article/S0022-5223(19)31676-9/fulltext.
        Figure thumbnail fx3
        Video 2Secondary (internal) flow path. With an open static circuit, flow is induced by the pump. Contrast injection flows selectively through the internal flow path, with dispersion of outflow at the equatorial midline. Internal path flow is essential to pump function for heat dissipation, for bearing flushing, and to reduce risk of recirculation, stasis, and thrombosis. Video available at: https://www.jtcvs.org/article/S0022-5223(19)31676-9/fulltext.
      • 9.
        No barrier to recirculation: to reduce thrombogenicity and obstruction risk.
      • 10.
        Small size: in situ implantation to avoid physical encroachment on adjacent structures (eg, aorta, common atrium, pulmonary veins).
      TCPC, Total cavopulmonary connection; SVC, superior vena cava; IVC, inferior vena cava; RPA, right pulmonary artery; LPA, left pulmonary artery.
      Although a von Karman viscous pump is a relatively weak pump, its inherent performance is ideal for the Fontan circulation. As a dynamic pump, it can provide consistent pressure rise across a wide range of cardiac output, making it responsive to physiologic demand. Low hydraulic efficiency also makes it less likely to generate suction, vein collapse, and cavitation and makes it tolerant of physiologic variation in systemic venous return. Similarly, it is unlikely to generate excessive downstream pressure, reducing the risk of lung perfusion injury. Most existing MCS devices are displacement pumps that must vary rotational speed to vary flow rate, making them susceptible to suction and unresponsive to physiologic demand, thereby limiting exercise capacity.
      The TCPC is the best-known geometry for passive cavopulmonary flow.
      • Bove E.L.
      • de Leval M.R.
      • Migliavacca F.
      • Balossino R.
      • Dubini G.
      Toward optimal hemodynamics: computer modeling of the Fontan circuit.
      Our strategy has therefore been to retain the TCPC as a default flow path and design an anatomically specific pump to add pressure energy within it. A central diverting body at the TCPC intersection will split incoming flow toward each outlet, significantly reducing turbulent energy loss.
      • Soerensen D.D.
      • Pekkan K.
      • de Zelicourt D.
      • Sharma S.
      • Kanter K.
      • Fogel M.
      • et al.
      Introduction of a new optimized total cavopulmonary connection.
      • Desai K.
      • Haggerty C.M.
      • Kanter K.R.
      • Rossignac J.
      • Spray T.L.
      • Fogel M.A.
      • et al.
      Haemodynamic comparison of a novel flow-divider Optiflo geometry and a traditional cavopulmonary connection.
      By rotating a central stabilizing body, suspended within the TCPC independent of the vessel walls, fluid pressure and velocity are increased, transforming it from a static flow diverter to a pump. We also anticipate that a permanently implanted device will eventually fail. In a Fontan circulation, a pump situated in the center of the flow path cannot impede flow (Figure 5). This is a critical safety consideration in young patients with Fontan circulations who have decades of life expectancy ahead of them.
      Figure thumbnail gr5
      Figure 5Pump implantation (conceptual). The cavopulmonary assist device is implanted in situ in the total cavopulmonary connection and serves to augment cavopulmonary flow modestly. The device is compatible with either lateral tunnel or extracardiac conduit Fontan construction. Graft extensions allow for suturing, with a total of 4 anastomoses. Grafts can be extended or tailored to address anatomic issues. To facilitate device exchange in the event of dysfunction, it is possible to replace the device through the modular “quick connects” that attach each limb to the pump body. Two pumps are depicted. The systemic ventricle performs most of the workload. It is served by a low-input cavopulmonary assist device to normalize preload to the systemic ventricle, in addition to reducing systemic venous pressure. SVC, Superior vena cava; PA, pulmonary artery; IVC, inferior vena cava.
      The ideal pressure rise for a cavopulmonary assist device is not yet clinically determined. We have shown in this study, and in other in vitro mock circulatory studies of a Fontan circulation, that 6 mm Hg pressure step-up will restore biventricular equivalency under normal physiologic conditions.
      • Rodefeld M.D.
      • Coats B.
      • Fisher T.
      • Giridharan G.A.
      • Chen J.
      • Brown J.W.
      • et al.
      Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
      • Rodefeld M.D.
      • Frankel S.H.
      • Giridharan G.A.
      Cavopulmonary assist: (Em)powering the univentricular Fontan circulation.
      This pressure rise correlates with normal circulatory physiology. The presence of increased pulmonary vascular resistance in the Fontan circulation may affect the magnitude of pressure rise required. Extremely elevated pulmonary resistance is incompatible with late Fontan survival; we therefore believe that the need for a pressure increase significantly greater than approximately 15 mm Hg is unlikely. Further, in the cavopulmonary circulation, the margin to apply an aggressive pressure step-up is limited by baseline systemic venous pressure. For example, the application of a 20 mm Hg pressure step-up in the TCPC in which baseline systemic venous pressure is 16 mm Hg will induce negative (−4 mm Hg) systemic venous pressure, leading to suction collapse and flow disruption.
      In general, MCS devices are not placed in line or in series with the circulation, but rather are placed in parallel (ie, paracardiac) because of the risk of device obstruction of the native flow path. This concern is amplified for a Fontan pump situated in the lowest pressure segment of the circulation, where seemingly trivial obstruction can induce hemodynamic instability. An implantable Fontan pump must have no potential for mechanical obstruction in the event of dysfunction.

      Emerging Technologies

      Concepts for dedicated Fontan circulatory support are currently evolving.
      • Rodefeld M.D.
      • Frankel S.H.
      • Giridharan G.A.
      Cavopulmonary assist: (Em)powering the univentricular Fontan circulation.
      • Broda C.R.
      • Taylor D.A.
      • Adachi I.
      Progress in experimental and clinical subpulmonary assistance for Fontan circulation.
      To date, nearly all have focused on modification of existing technology (unidirectional axial or centrifugal ventricular assist devices) to operate in the low-pressure Fontan environment. Although these devices may suffice as bridge therapy, they lack the technologic rigor required for safe and durable long-term support. They are incapable of supporting flow in all 4 limbs of the cavopulmonary connection and will induce back-pressure elevation in the opposing systemic venous territory. They are capable of mechanical obstruction, a prohibitive issue when the right-sided circulation is dependent on the device. Because of motor configuration, they may be bulky and not fit within the confines of the TCPC. Finally, if the Fontan pathway is modified to accommodate a unidirectional flow device, the resulting pathway will not support passive cavopulmonary flow in the event of device dysfunction. With respect to safety and durability, a Fontan pump that offers multidirectional flow, in situ anatomic placement, and a contingency plan for failure is preferable.

      Clinical Implications

      Cavopulmonary assist reframes single-ventricle palliation from a problem with no solution to a problem with an obvious solution. The challenge lies in creating a technology that can safely and durably implement it. If successfully translated, it would represent a first-ever targeted therapy for Fontan failure. It would also permit biventricular equivalency to be maintained in a univentricular circulation before clinical failure, preempting Fontan disease. The ideal timing for device intervention is not yet known. For patients with Fontan circulations, who have few options to alter their inevitable disease progression, this concept may fill a significant gap in the existing armamentarium of surgical therapies with a more comprehensive solution that has exciting potential to shift the paradigm for single-ventricle therapy from palliation to cure.
      The clinical vision for a Fontan-specific cavopulmonary assist device is greatly different from current MCS strategy. Rather than reserving MCS for end-stage salvage, it may be better to apply it preemptively for biventricular health maintenance. The device thus may not necessarily be limited to advanced Fontan failure, and may be considered earlier. For patients with Fontan circulations, the implications are starkly different: a lifetime of chronic disease progression with systemic MCS as end-stage therapy versus cavopulmonary assist to enable long-term biventricular health and wellness, potentially for decades.

      Limitations

      This report presents early-stage electromechanical feasibility for an implantable cavopulmonary assist device concept. Such a therapy has never been implemented clinically; discussion of clinical application and duration of support is therefore theoretic. Substantial technologic development remains to be performed. Thrombogenicity and hemolysis performance remain to be determined, and device durability is unknown. The power source has not yet been determined, although the device is envisioned to be powered wirelessly. Until a device is fully developed and enters clinical trial, the translational effect of long-term Fontan reversal will remain speculative.

      Conclusions

      After Fontan repair of single functional ventricle, patients are trapped for the remainder of their lives in a cycle of chronic circulatory inefficiency. Despite dramatic improvements in surgical technique, perioperative care, and long-term surveillance, late Fontan failure and attrition remain an intractable problem for which there is no primary or preventive therapy. An early-stage implantable cavopulmonary assist device shows promising potential to reverse the Fontan paradox and preempt Fontan-associated disease as a curative therapy. It represents a significant opportunity to shift the paradigm for single-ventricle care fundamentally to one based on biventricular health.

      Webcast

      Conflict of Interest Statement

      Dr Rodefeld discloses patent US 9,827,357 for cavopulmonary viscous impeller assist device and method. Dr Giridharan discloses interuniversity contract with IU for device (Dr Rodefeld) and consultant status for NuPulseCV, which is developing a counterpulsation device; the present work is unrelated (Fontan support). All other authors have nothing to disclose with regard to commercial support.

      Supplementary Data

      References

        • Fontan F.
        • Baudet E.
        Surgical repair of tricuspid atresia.
        Thorax. 1971; 26: 240-248
        • Kreutzer G.
        • Galindez E.
        • Bono H.
        • DePalma C.
        • Laura J.P.
        An operation for the correction of tricuspid atresia.
        J Thorac Cardiovasc Surg. 1973; 66: 613-621
        • de Leval M.R.
        The Fontan circulation: what have we learned? What to expect?.
        Pediatr Cardiol. 1998; 19: 316-320
        • Rychik J.
        The relentless effects of the Fontan paradox.
        Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2016; 19: 37-43
        • Gewillig M.
        • Brown S.C.
        The Fontan circulation after 45 years: update in physiology.
        Heart. 2016; 102: 1081-1086
        • Rodefeld M.D.
        • Boyd J.H.
        • Myers C.D.
        • Lalone B.J.
        • Bezruczko A.
        • Potter A.W.
        • et al.
        Cavopulmonary assist: circulatory support for the univentricular Fontan circulation.
        Ann Thorac Surg. 2003; 76: 1911-1916
        • Rodefeld M.D.
        • Coats B.
        • Fisher T.
        • Giridharan G.A.
        • Chen J.
        • Brown J.W.
        • et al.
        Cavopulmonary assist for the univentricular Fontan circulation: von Karman viscous impeller pump.
        J Thorac Cardiovasc Surg. 2010; 140: 529-536
        • Fontan F.
        • Kirklin J.W.
        • Fernandez G.
        • Costa F.
        • Naftel D.C.
        • Tritto F.
        • et al.
        Outcome after a “perfect” Fontan operation.
        Circulation. 1990; 81: 1520-1536
        • Kreutzer C.
        • Kreutzer J.
        • Kreutzer G.O.
        Reflections on five decades of the Fontan Kreutzer procedure.
        Front Pediatr. 2013; 1: 45-60
        • de Leval M.R.
        The Fontan circulation: a challenge to William Harvey?.
        Nat Clin Pract Cardiovasc Med. 2005; 2: 202-208
        • Rychik J.
        Forty years after the Fontan: a failed strategy.
        Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2010; 13: 96-100
        • Veldtman G.R.
        • Opotowsky A.R.
        • Wittekind S.G.
        • Rychik J.
        • Penny D.J.
        • Fogel M.
        • et al.
        Cardiovascular adaptation to the Fontan circulation.
        Congenit Heart Dis. 2017; 12: 699-710
        • Diller G.P.
        • Kempny A.
        • Alonso-Gonzalez R.
        • Swan L.
        • Uebing A.
        • Li W.
        • et al.
        Survival prospects and circumstances of death in contemporary adult congenital heart disease patients under follow-up at a large tertiary centre.
        Circulation. 2015; 132: 2118-2125
        • Pundi K.N.
        • Johnson J.N.
        • Dearani J.A.
        • Li Z.
        • Hinck C.A.
        • Dahl S.H.
        • et al.
        40-year follow-up after the Fontan operation: long-term outcomes of 1,052 patients.
        J Am Coll Cardiol. 2015; 66: 1700-1710
        • Poh C.L.
        • Zannino D.
        • Weintraub R.G.
        • Winlaw D.S.
        • Grigg L.E.
        • Cordina R.
        • et al.
        Three decades later: the fate of the population of patients who underwent the atriopulmonary Fontan procedure.
        Int J Cardiol. 2017; 231: 99-104
        • Rodefeld M.D.
        The Fontan circulation: time for a moon shot?.
        J Thorac Cardiovasc Surg. 2018; 156: 252-253
        • Anderson P.A.
        • Sleeper L.A.
        • Mahony L.
        • Colan S.D.
        • Atz A.M.
        • Breitbart R.E.
        • et al.
        Contemporary outcomes after the Fontan procedure: a pediatric heart network multicenter study.
        J Am Coll Cardiol. 2008; 52: 85-98
        • Averin K.
        • Hirsch R.
        • Seckeler M.D.
        • Whiteside W.
        • Beekman R.H.
        • Goldstein B.H.
        Diagnosis of occult diastolic dysfunction late after the Fontan procedure using a rapid volume expansion technique.
        Heart. 2016; 102: 1109-1114
        • De Mey W.
        • Cools B.
        • Heying R.
        • Budts W.
        • Louw J.J.
        • Boshoff D.E.
        • et al.
        Can a volume challenge pinpoint the limiting factor in a Fontan circulation?.
        Acta Cardiol. 2015; 70: 536-542
        • Rathod R.H.
        • Prakash A.
        • Powell A.J.
        • Geva T.
        Myocardial fibrosis identified by cardiac magnetic resonance late gadolinium enhancement is associated with adverse ventricular mechanics and ventricular tachycardia late after Fontan operation.
        J Am Coll Cardiol. 2010; 55: 1721-1728
        • Ross H.J.
        • Law Y.
        • Book W.M.
        • Broberg C.S.
        • Burchill L.
        • Cecchin F.
        • et al.
        • American Heart Association Adults With Congenital Heart Disease Committee of the Council on Clinical Cardiology and Council on Cardiovascular Disease in the Young, the Council on Cardiovascular Radiology and Intervention, and the Council on Functional Genomics and Translational Biology
        Transplantation and mechanical circulatory support in congenital heart disease: a scientific statement from the American Heart Association.
        Circulation. 2016; 133: 802-820
        • Bove E.L.
        • de Leval M.R.
        • Migliavacca F.
        • Balossino R.
        • Dubini G.
        Toward optimal hemodynamics: computer modeling of the Fontan circuit.
        Pediatr Cardiol. 2007; 28: 477-481
        • Sundareswaran K.S.
        • Pekkan K.
        • Dasi L.P.
        • Whitehead K.
        • Sharma S.
        • Kanter K.R.
        • et al.
        The total cavopulmonary connection resistance: a significant impact on single ventricle hemodynamics at rest and exercise.
        Am J Physiol Heart Circ Physiol. 2008; 295: H2427-H2435
        • Tang E.
        • Restrepo M.
        • Haggerty C.M.
        • Mirabella L.
        • Bethel J.
        • Whitehead K.K.
        • et al.
        Geometric characterization of patient-specific total cavopulmonary connections and its relationship to hemodynamics.
        JACC Cardiovasc Imaging. 2014; 7: 215-224
        • Marsden A.L.
        • Bernstein A.J.
        • Reddy V.M.
        • Shadden S.C.
        • Spilker R.L.
        • Chan F.P.
        • et al.
        Evaluation of a novel Y-shaped extracardiac Fontan baffle using computational fluid dynamics.
        J Thorac Cardiovasc Surg. 2009; 137: 394-403
        • Arnaoutakis G.J.
        • Blitzer D.
        • Fuller S.
        • Eckhauser A.W.
        • Montenegro L.M.
        • Rossano J.W.
        • et al.
        Mechanical circulatory support as bridge to transplantation for the failing single ventricle.
        Ann Thorac Surg. 2017; 103: 193-197
        • Horne D.
        • Conway J.
        • Rebeyka I.M.
        • Buchholz H.
        Mechanical circulatory support in univentricular hearts: current management.
        Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2015; 18: 17-24
        • Adachi I.
        • Burki S.
        • Fraser Jr., C.D.
        Current status of pediatric ventricular assist device support.
        Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2017; 20: 2-8
        • Woods R.K.
        • Ghanayem N.S.
        • Mitchell M.E.
        • Kindel S.
        • Niebler R.A.
        Mechanical circulatory support of the Fontan patient.
        Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2017; 20: 20-27
        • Prêtre R.
        • Häussler A.
        • Bettex D.
        • Genoni M.
        Right-sided univentricular cardiac assistance in a failing Fontan circulation.
        Ann Thorac Surg. 2008; 86: 1018-1020
        • Furey S.A.
        • Zieske H.A.
        • Levy M.N.
        The essential function of the right ventricle.
        Am Heart J. 1984; 107: 404-410
        • Soerensen D.D.
        • Pekkan K.
        • de Zelicourt D.
        • Sharma S.
        • Kanter K.
        • Fogel M.
        • et al.
        Introduction of a new optimized total cavopulmonary connection.
        Ann Thorac Surg. 2007; 83: 2182-2190
        • Desai K.
        • Haggerty C.M.
        • Kanter K.R.
        • Rossignac J.
        • Spray T.L.
        • Fogel M.A.
        • et al.
        Haemodynamic comparison of a novel flow-divider Optiflo geometry and a traditional cavopulmonary connection.
        Interact Cardiovasc Thorac Surg. 2013; 17: 1-7
        • Rodefeld M.D.
        • Frankel S.H.
        • Giridharan G.A.
        Cavopulmonary assist: (Em)powering the univentricular Fontan circulation.
        Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2011; 14: 45-54
        • Broda C.R.
        • Taylor D.A.
        • Adachi I.
        Progress in experimental and clinical subpulmonary assistance for Fontan circulation.
        J Thorac Cardiovasc Surg. 2018; 156: 1949-1956

      Linked Article

      • Commentary: Subpulmonary pump in Fontan circulation: Can it be a breakthrough in treatment of the single ventricle?
        The Journal of Thoracic and Cardiovascular SurgeryVol. 158Issue 6
        • Preview
          It has been almost half a century since the first successful report of Fontan surgery.1 Without any doubt, this was one of the major breakthroughs in the history of cardiac surgery and brought a ray of light to a previously unrepairable and lethal heart condition. However, long-term survival after the Fontan operation is still far from satisfactory.2 Efforts to improve survival to date have focused on optimization of the design of the Fontan circuit per se to make the best use of the single energy source from the single ventricle.
        • Full-Text
        • PDF
        Open Archive
      • Commentary: Cavopulmonary assist: Closing in on the white whale of single-ventricle palliation
        The Journal of Thoracic and Cardiovascular SurgeryVol. 158Issue 6
        • Preview
          Improving the long-term survival and quality of life for patients with single-ventricle heart disease remains the single greatest unmet need in pediatric heart surgery and pediatric cardiology. Although the surgical outcomes of single-ventricle palliation have been celebrated and have steadily improved over time, the unsightly truth is that the Fontan circulation is a grossly abnormal circulatory configuration, which leads to a myriad of progressive, debilitating, and ultimately fatal medical problems that accumulate over time.
        • Full-Text
        • PDF
        Open Archive
      • Commentary: The future fourth stage of single-ventricle palliation
        The Journal of Thoracic and Cardiovascular SurgeryVol. 158Issue 6
        • Preview
          Five decades after Francis Fontan first described his procedure for palliation of tricuspid atresia, modified versions have enabled long-term survival of a broad array of patients with univentricular circulation.1 It has long been recognized, however, that the sequelae of the total cavopulmonary connection—including systemic venous congestion, hepatic dysfunction, renal insufficiency, changes in pulmonary vascular physiology, and elevated systemic afterload—result in long-term failure of even the “perfect” Fontan circulation.
        • Full-Text
        • PDF
        Open Archive