We examine the implications of the dramatic controversial technique of valve cracking to optimize results after the VIV procedure that has been recently reported.
Use of bioprosthetic surgical heart valves (SHVs) is on the rise, especially in younger patients.
A large number of these patients will likely require at least 1 reintervention during their lifetime, and valve-in-valve (VIV) has emerged as a novel option whereby a transcatheter heart valve (THV) is implanted within a failing SHV. VIV has recently been approved in patients deemed high or extreme risk for reoperative surgical aortic valve replacement (SAVR).
Transcatheter tricuspid valve-in-valve implantation for the treatment of dysfunctional surgical bioprosthetic valves: an international, multicenter registry study.
Relationships among conduit type, pre-stenting, and outcomes in patients undergoing transcatheter pulmonary valve replacement in the prospective North American and European melody valve trials.
Potential advantages of VIV therapy over a high-risk reoperation include a shorter and simpler procedure, lower mortality, reduced risk of morbidities, including bleeding and renal injury, shorter length of stay, and more rapid recovery.
Transcatheter tricuspid valve-in-valve implantation for the treatment of dysfunctional surgical bioprosthetic valves: an international, multicenter registry study.
Relationships among conduit type, pre-stenting, and outcomes in patients undergoing transcatheter pulmonary valve replacement in the prospective North American and European melody valve trials.
defined “10 Commandments” to describe a perfect valve substitute, but current mechanical and bioprosthetic SHVs fall short of those features. Mechanical valves can be durable but require lifelong anticoagulation, whereas bioprosthetic valves may not need anticoagulation and are noise free but will definitely degenerate.
In a way, VIV can take us a step closer to achieving these 10 Commandments if we can demonstrate the safety and efficacy of this approach, and in addition modify design of the bioprosthetic SHV to accommodate VIV, that is, THV proof designs.
However, VIV therapy in degenerated SHVs has several potential pitfalls, which need to be overcome and its effectiveness proven if the therapy were to be offered to lower-risk patients. These are the risks of malposition, coronary obstruction, leaflet thickening/thrombosis, and “Russian doll” effect.
Transcatheter Heart Valve Malposition
Malposition is a well-recognized complication whereby THV is placed too low or too high, resulting in suboptimal hemodynamics or embolization. Other than operator experience, the main cause is suboptimal fluoroscopic landmarks in certain SHV. Because the VIV procedure is performed under fluoroscopy, an SHV with no or poor fluoroscopic landmarks poses a challenge in optimal THV positioning. This is confirmed by the fact that stentless valves and Mosaic bioprostheses (Medtronic Inc, Galway, Ireland) have the highest incidence of malposition.
Coronary obstruction is a life-threatening complication after the VIV procedure. During a VIV procedure, the coronary arteries are at risk of occlusion as the stent posts and leaflets of the SHV are pushed outward toward the coronary ostia. Risk of coronary obstruction after a VIV procedure is higher than after transcatheter aortic valve replacement (TAVR) in a native aortic valve, with an incidence of 2.5% to 3.5% compared with less than 1%.
Computed tomography assessment for transcatheter aortic valve in valve implantation: the Vancouver approach to predict anatomical risk for coronary obstruction and other considerations.
It is a combination of small aortic root, relationship of the coronary ostia to the SHV, and certain valve designs (ie, SHV with leaflets mounted outside the stent frame) that increase coronary obstruction risk.
Computed tomography assessment for transcatheter aortic valve in valve implantation: the Vancouver approach to predict anatomical risk for coronary obstruction and other considerations.
If reoperation is not deemed feasible, VIV can be performed with coronary protection and immediate PCI if coronary flow becomes jeopardized. A novel technique named “Bioprosthetic Aortic Scallop Intentional Laceration to prevent Iatrogenic Coronary Artery obstruction” (BASILICA) may also reduce the risk of coronary obstruction during VIV TAVR, especially if the aortic root to coronary anatomy is not favorable for percutaneous coronary intervention.
The third issue is related to the extent of THV oversizing relative to the implanted bioprosthesis, especially in small SHVs. VIV is susceptible to leaflet thickening or thrombosis, especially in smaller-sized SHVs.
Several mechanisms may be responsible, including insufficient eddy currents to achieve optimal leaflet closing, increased blood stasis on THV leaflets and adjacent frame inside the surgical valve, and under-expansion of the THV inside the small SHV, leading to crowding and a pinwheel effect on the THV leaflets.
Matched comparison of self-expanding transcatheter heart valves for the treatment of failed aortic surgical bioprosthesis: insights from the valve-in-valve international data registry (VIVID).
Transcatheter replacement of failed bioprosthetic valves: large multicenter assessment of the effect of implantation depth on hemodynamics after aortic valve-in-valve.
Although some factors to optimize the hemodynamics can be controlled by the operator, such as the type of THV used and ensuring optimal depth of implantation,
Matched comparison of self-expanding transcatheter heart valves for the treatment of failed aortic surgical bioprosthesis: insights from the valve-in-valve international data registry (VIVID).
Transcatheter replacement of failed bioprosthetic valves: large multicenter assessment of the effect of implantation depth on hemodynamics after aortic valve-in-valve.
it is clear that there are certain limitations set by the SHV in situ, especially when moderate to severe PPM already exists in the SHV. Each SHV is unique in its construct but has 3 components: sewing ring, stent frame, and leaflets.
Effect of valve design on the stent internal diameter of a bioprosthetic valve: a concept of true internal diameter and its implications for the valve-in-valve procedure.
Stent posts and leaflets are deflected outward, but the inflow dimensions at the level of “neo-annulus” remain constant; thus, the THV remain most constrained at this level (Figure 1). Until recently, it was thought to be impossible to expand the neo-annulus.
Figure 1Concept of neo-annulus. A, Image of Mitroflow valve (Sorin Group, Arvada, Colo). B, Fluoroscopic appearance of the Mitroflow valve sewing ring is visible. C, Balloon inflation within the Mitroflow valve demonstrates zone of restrain (black arrows). D, SAPIEN XT (Edwards Lifesciences, Irvine, Calif) implantation within Mitroflow demonstrates uneven expansion of the SAPIEN XT valve with a “waist” at the level of the sewing ring (black arrows).
; they fractured a Mitroflow SHV (Sorin Group, Arvada, Colo) with a high-pressure balloon, so-called balloon valve fracture (BVF). The design features of the Mitroflow device, in which the stent frame made of polymer is attached by a joint and yields to the excessive force applied from within, allows room for expansion. Subsequent bench testing using different THV designs confirmed improvements in gradient and effective orifice area after BVF.
Valve-in-valve transcatheter aortic valve replacement and bioprosthetic valve fracture comparing different transcatheter heart valve designs: an ex vivo bench study.
Additional reports have demonstrated early feasibility of intentionally fracturing the sewing ring of SHV to improve the expansion of THV implanted in small SHVs.
Few critical observations were made during bench testing. The size of the balloon required to fracture the SHV was at least equal to the label size. The balloon had to be a noncompliant, high-pressure balloon made of material such as Kevlar. Valve cracking occurred typically at high pressure, that is, 25 to 30 atm. Valve cracking appeared to have not disrupted the fabric covering of the SHV. Among the SHV types implanted in the last 2 decades, the Mitroflow, Magna (Edwards Lifesciences, Irvine, Calif), Magna Ease (Edwards Lifesciences), Mosaic (Medtronic Inc, Minneapolis, Minn), and Biocor Epic (Abbott, St Paul, Minn) surgical valves could be successfully fractured using noncompliant high-pressure angioplasty balloons, whereas the Hancock II (Medtronic Inc) and Trifecta (Abbott) could not.
The Trifecta SHV has a titanium frame without a joint and cannot be fractured. The construct of the sewing band in the Hancock II also resists fracturing.
Although in vitro testing has demonstrated that the sewing cuff was not disrupted despite fracture of the internal valve frame, the main concern in prescribing this approach routinely is the risk of annular injury and rupture. This complication has been anecdotally described but not reported. Chhatriwalla and colleagues
have reported on 20 consecutive clinical cases of BVF performed in 9 centers across the United States as part of an ongoing study. A variety of SHVs were treated, including the Perimount (Carpentier Edwards, Irvine, Calif), Magna, Carpentier Edwards porcine, Mosaic, Mitroflow, Epic (Abbott), and Epic Supra (Abbott). Label size ranged from 19 to 25 mm, but only SHVs with a true internal diameter less than 21 mm were treated. Types of THV used (ie, SAPIEN XT, SAPIEN 3 [Edwards Lifesciences], or CoreValve Evolut [Medtronic Inc]) and timing of the balloon fracture (ie, before or after implantation of the THV) were left to the discretion of the operator. Although the burst pressure was significant highly than what we would encounter during valve implantation with balloon-expandable valves (ie, from 8 to 24 atm), no intraprocedural complications were reported. Improved hemodynamics and frame expansion of THV implanted in fractured SHVs were observed.
When device failure translates to therapeutic efficacy: intentional fracture of bioprosthetic valve rings as an adjunctive technique in transcatheter valve-in-valve replacement.
This novel technique may potentially improve PPM and hemodynamic outcomes in patients with small surgical bioprostheses undergoing VIV TAVR.
There remain many unknowns of balloon fracturing of the stent frame of SHV in the VIV procedure. Sizing the appropriate balloon for the particular surgical valve has not been established.
A larger balloon may fracture the SHV at a lower inflation pressure, whereas a smaller balloon may generate lower radial force against the basal ring of the SHV. Using a larger balloon may risk injuring adjacent aortic root structures, such as the sinotubular junction or left ventricular outflow tract. Improved expansion of the THV may further displace the thickened SHV leaflets outward, increasing the risk of coronary obstruction, because the SHV leaflet-to-coronary distance will be reduced as the result of THV expansion in an unpredictable manner. The BASILICA technique may mitigate against this potential risk but is currently complex and should be performed only at experienced centers. The timing of balloon fracturing is also unclear. Fracturing before VIV implantation may cause leaflet damage and embolization, or acute hemodynamic decompensation, especially in patients without significant prosthetic aortic regurgitation. On the other hand, balloon fracturing after VIV implantation may damage the leaflets of the THV, given that larger-sized noncompliant balloons will be used under higher bursting pressure than what would be used typically postdilatation. Acute THV failure as a result of surgical valve fracturing has been anecdotally described and may lead to potentially catastrophic complications. Also, it remains unclear whether after BVF a THV can become fully expanded and whether a more expanded THV may reduce the risk of leaflet thrombosis. Furthermore, given that BVF was intended to address the concern of residual elevated THV gradient after VIV TAVR, especially in small SHVs, we do not believe BVF would be necessary if the post–VIV TAVR mean gradient is less than 20 mm Hg or mirrors expected SHV gradients. Using both echocardiography and invasive hemodynamics to assess THV gradient at the time of the procedure would be helpful to determine whether BVF would be necessary. Judicious balloon use after dilatation with a balloon size equivalent of 1 mm greater than the internal diameter of SHV may be sufficient to reduce the THV gradient. This more conservative strategy may be especially relevant in SHV, in which the primary dysfunction is prosthetic aortic stenosis and the SHV leaflets would be stiff and calcified, and there would be a risk of coronary obstruction if BVF were to be performed. Finally, significant PPM and the “Russian doll” effect may remain in small SHVs despite BVF. Although the THV effective orifice area and gradient may improve after BVF, preexisting severe PPM with the small SHV may remain, and the potential benefit of BVF and VIV TAVR may be attenuated. Reoperative SAVR, with possible aortic root enlargement in patients with small annuli who are not high risk for surgery, may be a reasonable alternative, given the outcomes in reoperative SAVR remained acceptable and were found to be similar to those of VIV TAVR.
Thus, we believe that further bench testing to understand the implication of this novel therapy is a must and should be used only in select groups of high- or extreme-risk patients.
Conclusions
Going a step further, it is obvious that for VIV therapy to be a viable option in patients with a degenerated SHV, the SHV design must be modified to minimize or eliminate the aforementioned limitations and complications. We have already witnessed this in the form of the Crown valve (next-generation Mitroflow) manufactured by LivaNova (London, England) with sewing ring indents to help identify position of the stent posts under fluoroscopy (Figure 2, A and B), the Inspiris valve manufactured by Edwards Lifesciences with label size markers and expansion mechanism in its basal ring (Figure 2, C and D), and the Avalus valve manufactured by Medtronic Inc, with a design to minimize the risk of coronary obstruction. Such design changes in SHV should facilitate future VIV procedures, but until then, ingenious methods such as SHV fracturing may be the only way to optimize the VIV procedure in select patients.
Figure 2TAVR-friendly SHV designs. A, Mitroflow SHV image and fluoroscopic appearance. B, Crown SHV with cut-outs in the sewing ring at the level of stent posts (black arrows). C, Magna Ease (Edwards Lifesciences): image and fluoroscopic appearance. D, Inspiris SHV (Edwards Lifesciences) with label size markers and expansion mechanism in the basal band (white arrow).
Transcatheter tricuspid valve-in-valve implantation for the treatment of dysfunctional surgical bioprosthetic valves: an international, multicenter registry study.
Relationships among conduit type, pre-stenting, and outcomes in patients undergoing transcatheter pulmonary valve replacement in the prospective North American and European melody valve trials.
Computed tomography assessment for transcatheter aortic valve in valve implantation: the Vancouver approach to predict anatomical risk for coronary obstruction and other considerations.
Matched comparison of self-expanding transcatheter heart valves for the treatment of failed aortic surgical bioprosthesis: insights from the valve-in-valve international data registry (VIVID).
Transcatheter replacement of failed bioprosthetic valves: large multicenter assessment of the effect of implantation depth on hemodynamics after aortic valve-in-valve.
Effect of valve design on the stent internal diameter of a bioprosthetic valve: a concept of true internal diameter and its implications for the valve-in-valve procedure.
Valve-in-valve transcatheter aortic valve replacement and bioprosthetic valve fracture comparing different transcatheter heart valve designs: an ex vivo bench study.
When device failure translates to therapeutic efficacy: intentional fracture of bioprosthetic valve rings as an adjunctive technique in transcatheter valve-in-valve replacement.